ENG

    CAMAG® Automatic Developing Chamber 2 (ADC 2)

    The ADC 2 offers convenience, safety, and reproducibility for isocratic developments of  20 x 10 cm HPTLC glass plates in full compliance with Ph. Eur. Chapter 2.8.25 and USP Chapter <203>.

    The ADC 2 is a device for reproducible plate development. It performs the development step fully automated, and independent of environmental effects. The activity and pre-conditioning of the layer, chamber saturation, developing distance and final drying can be preset and automatically monitored by the ADC 2. Two modes of operation are possible: stand-alone with input of parameters via keypad, or remote operation from visionCATS with process monitoring, documentation of operating parameters, and reporting.

    With visionCATS the ADC 2 can be IQ/OQ qualified and used in a cGMP environment. Operated with the mentioned software, the ADC 2 supports compliance with 21 CFR Part 11.

    Key Features

    • Fully automated development
    • Development in 20 x 10 cm Twin Trough Chamber
    • HPTLC plates (20 x 10 cm)
    • Operation in stand-alone mode or software-controlled
    • Fully traceable operation
    • Option “Humidity Control” allows reproducible chromatography

    Technical Specifications

    Link

    Video

    Electrical Connections: 100–240 V AC 50/60Hz, 20W

    Dimensions (W x D x H): 330 x 330 x 520 mm

    Weight: 18.5 kg

    For more information please visit: camag adc 2

    Privacy Overview

    This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.