CAMAG® Linomat 5
With the Linomat 5 samples are sprayed onto TLC/HPTLC plates in the form of bands with nitrogen or compressed air. Application is automatic, only changing the syringe (filling, inserting and rinsing) is manual. The Linomat 5 is suitable for small sample throughput.
Key Features
- Semi-automatic application
- Application of spots or bands
- Any plate format up to 20 x 20 cm
- Spray-on application
- Data input through visionCATS software or keypad
Software-controlled operation with visionCATS
The Linomat 5 is controlled by visionCATS, the new generation of HPTLC software. Instrument handling is now easier and more convenient. Simply select a plate format, fill in the sequence table and start working. With visionCATS, the Linomat 5 can be IQ/OQ qualified and used in a cGMP environment.
Operation in stand-alone mode
For those who use the Linomat 5 infrequently stand-alone mode is available. Up to ten methods can be entered either manually via the keypad or downloaded to the instrument from a computer running a licensed visionCATS program. In stand-alone mode the keypad is used to enter application parameters or to select a saved method.
Technical Specifications
Link
Object support: For objects up to 20 x 20 cm
Stage drive: Stepping motor 3200 steps/rotation, 8 steps = 0.1 mm
Band length 0 (spot) – 190 mm in steps of 0.1 mm
Speed approx. 10 mm/s; the speed is automatically adapted such that a whole number of passages is reached with the volume selected.
Dosage syringe drive:
Stepping motor 1600 steps/rotation
100 nL = 120 steps with 100 µL syringe
100 nL = 24 steps with 500 µL syringe
Sample dosage syringe: selectable 100 µL or 500 µL
Memory: 10 methods, backup min. 10 years
LCD display: 2 lines of 16 characters ea.
Power connection: 85–250 V~ 47–63 Hz 30 VA
Gas supply: 4–6 bar (60–90 PSI) preferably nitrogen; consumption approx. 1.0 L/min
Dimensions (W x D x H): 360 x 510 x 410 mm
Weight: 12.5 kg
For more information visit camag linomat 5