ENG

    CONNECT LT

    Adrona Connect LT is water polishing system which requires a connection to a water pre-treatment unit or a distilled water distribution system.

    It produces ultrapure water that complies with ISO 3696 Grade 1 water requirements.

    The new version of market approved laboratory water polishing system – Connect received totally new electronics and software. As a result, there are number of the new features – full color touch screen, USB and Ethernet Connections and available history of dispense reports.

    Connect” system can be used in two configurations: pressurized and non-pressurized feed water configuration.

    FEATURES

    SPECIFICATION

    CONSUMABLES

    KEY FEATURES:

    • Requires connection to a pre-treated water distribution loop
    • Compact – space saving design
    • Cost effective – large DI cartridge
    • Easy data export via USB-C
    • Convenient user interface
    • 7″ touchscreen
    APPLICATION
    LT
    LT BIO
    Reagent preparation
    *
    *
    Flame AAS
    *
    *
    Autoclaves
    *
    *
    ICP-MS & ICP-OES
    *
    *
    Ion chromatography
    *
    *
    Liquid chromatography (HPLC)
    *
    *
    Gas chromatography
    *
    *
    TOC measurements
    *
    *
    Flow cytometry
    *
    Cell and tissue culture
    *
    Molecular biology
    *

    FEED WATER REQUIREMENTS:

    • Feed water: Pre-treated water
    • Conductivity: <100 μS/cm
    • TOC: <50 ppb
    • Pressure: 0.4–4.0 bar
    • Temperature: 5–35 °C

    The feed water must be pre-treated by reverse osmosis, distillation, electrodeionisation or deionisation technologies.

    Parameters
    Connect LT
    Connect LT Bio
    Catalogue No.
    CB-1801
    CB-1803
    Grade 1 water resistivity at 25 °C
    18,2 MΩ x cm
    18,2 MΩ x cm
    Grade 1 water conductivity at 25 °C
    0,055 μS/cm
    0,055 μS/cm
    TOC level
    < 5 ppb*
    < 5 ppb*
    Rnase
    -
    < 1 pg/ml
    Dnase
    -
    < 5 pg/ml
    Bacteria
    < 0,01 CFU/ml
    < 0,01 CFU/ml
    Endotoxins
    < 0,15 EU/ml
    < 0,001 EU/ml
    Particles >0.22 μm
    < 1/ml
    < 0,05/ml
    Dispense rate, ultrapure water
    1,5–2 l/min
    1,5–2 l/min
    Dimensions (WxDxH), cm
    29x44x61 cm
    29x44x61 cm
    System weight, kg
    17 kg
    17 kg
    Operating weight, kg
    20 kg
    20 kg

    * In appropriate operating conditions <2 ppb, otherwise normally <5 ppb

    CONSUMABLES

    Part number
    Description
    Comments
    10031
    Polishing Q w/ quick connectors
    Depends on water consumption
    10033
    Polishing QC w/ quick connectors
    Depends on water consumption
    10037
    Polishing QH w/ quick connectors
    Depends on water consumption
    10018
    Replacement photooxidation UV bulb
    10013
    Replacement 0.22 μm dispense filter, sterile
    10120
    Replacement ultrafilter
    For Bio configuration
    Privacy Overview

    This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.