ENG

    Crystal EX

    The Adrona Crystal EX system is the economic model of a multi-purpose water purification system. Despite its affordable price, the Crystal EX system is capable of producing both Grade 1 and Grade 2 water. The system is based on reverse osmosis and ion exchange technologies optimized for the best quality. Crystal EX has an excellent price/performance ratio and very low operation costs.

    There are 3 configurations available: Crystal EX Trace, Crystal EX HPLC, Crystal EX Bio.

    FEATURES

    SPECIFICATION

    CONSUMABLES

    KEY FEATURES:

    • Reliable operation
    • Low running costs
    • Great price/performance
    • Easy installation and maintenance
    Parameters
    TRACE
    HPLC
    BIO
    Reagent preparation
    *
    *
    *
    Flame AAS
    *
    *
    *
    Autoclaves
    *
    *
    *
    ICP-MS
    *
    *
    *
    ICP-OES
    *
    *
    *
    Ion chromatography
    *
    *
    *
    Liquid chromatography
    *
    *
    Gas chromatography
    *
    *
    TOC measurements
    *
    *
    Flow cytometry
    *
    Cell and tissue culture
    *
    Molecular biology
    *

    FEED WATER REQUIREMENTS:

    • Feed water: Potable tap water
    • Pressure: 0.4–4.0 bar
    • Conductivity: <1500 μS/cm
    • Temperature: 5–35 °C

    If the feed water does not meet specifications, additional pre-treatment solutions are available.

    Parameters
    Trace
    HPLC
    Bio
    Grade 1 water resistivity at 25 °C
    18,2 MΩ x cm
    18,2 MΩ x cm
    18,2 MΩ x cm
    Grade 1 water resistivity at 25 °C
    0,055 μS/cm
    0,055 μS/cm
    0,055 μS/cm
    Grade 2 water resistivity at 25 °C
    >10 MΩ x cm
    >10 MΩ x cm
    >10 MΩ x cm
    Grade 2 water conductivity at 25 °C
    < 0,1 μS/cm
    < 0,1 μS/cm
    < 0,1 μS/cm
    TOC level
    < 10 ppb
    < 5 ppb*
    < 5 ppb*
    Rnase
    -
    -
    < 1 pg/ml
    Dnase
    -
    -
    < 5 pg/ml
    Bacteria
    < 0,01 CFU/ml
    < 0,01 CFU/ml
    < 0,01 CFU/ml
    Endotoxins
    < 0,15 EU/ml
    < 0,15 EU/ml
    < 0,001 EU/ml
    Particles >0.22 μm
    < 1/ml
    < 1/ml
    < 1/ml
    Nominal flow, pure water (to storage tank)
    10 l/h
    10 l/h
    10 l/h
    Volumetric dispense
    -
    -
    -
    Dimensions (WxDxH), cm
    40*35*55 cm
    40*35*55 cm
    40*35*55 cm
    System weight, kg
    17
    18
    19
    Operating weight, kg
    24
    25
    26

    *In appropriate operating conditions <2 ppb, otherwise normally <5 ppb.

    CONSUMABLES

    Part number
    Description
    Comments
    10320
    Replacement pre-filter, Crystal EX
    10310
    Replacement deionization module
    10030
    Replacement polishing module
    10017
    Sterilization UV bulb
    Bio configuration
    10018
    Replacement photooxidation UV bulb
    HPLC and Bio configuration
    10013
    Replacement 0.22 µm dispense microfilter, sterile
    Trace and HPLC configuration
    10120
    Replacement ultrafilter
    Bio configuration
    Privacy Overview

    This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.